UTRIX-16
- Durable metal USB connector - No separate Interface Cradle required
- Automatically generates a PDF report - No special software required
- Encrypted LTD file generated for a more detailed analysis
- Up to 16,129 recordings - provides over 3 months at 10 minute samples
- Suitable for any multi-use application
The LogTag® UTRIX-16 satisfies the growing need for a cost effective and reliable multi-trip USB temperature logger. It requires no special hardware or proprietary software to access the recorded data and can generate a fully detailed PDF report. LogTag PDF USB Data Recorders are the ultimate in paperless monitors.
The LogTag UTRIX-16 USB data recorder stores real-time temperature readings over a measurement range of -25°C to +70°C (-13°F to +158°F), displays user-configurable alerts, and comes with an integrated, long-life USB connector. The USB temperature logger is enclosed in the same robust and durable polycarbonate case common to other LogTag products.
Specifications: Model UTRIX-16
Recording/Operating/Recommended Storage Temperature Range -25°C to +70°C (-13°F to +158°F)
Resolution 0.1°C (0.1°F) across entire range
Accuracy
Better than ±0.5°C (±0.9°F) for -20°C to +40°C (-4°F to +104°F)
Better than ±0.7°C (±1.3°F) for -25°C to -20°C (-13°F to -4°F) & +40°C to +60°C (+104°F to +140°F)
Better than ±0.8°C (±1.5°F) for +60°C to +70°C (+140°F to +158°F)
Sensor Reaction Time Typically less than 7 minutes(T90) in moving air (1m/s) in EN12830:1999
Capacity 16,129 real time temperature recordings, non-volatile memory
Sampling Frequency Configurable from 30 seconds to 18 hours
Logging Start Options Push button start or specific date/time
Recording Indication Flashing 'OK' indicator / flashing 'ALERT' indicator.
Download Time Typically with full memory (16,129 readings) in less than 30 seconds from time of insertion to availability of PDF report.
Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured)
Environmental IEC 60529: IP64 with USB cap fitted. Fits into IP67 Protective Enclosure 200-000020
Power Source CR2450 3V non-rechargeable Lithium Manganese battery non-replaceable
Alarm Functions High/low, single, consecutive and cumulative alarm events, ‘OK’/’ALERT’ indicators
Connection Interface USB 2.0, A-type plug
Software PDF Reader, LogTag Analyzer® 2.6 or higher
Battery Life Typically 2 years with normal use (15 minute sampling interval, monthly download)
Size 93mm(H) x 54.5mm(W) x 8.6mm(T) including protective USB cap.
Weight 35g
Case Material Polycarbonate
EMC Compliance
EMC Compliance
Tested & complies with EC EMC directives (EN 50081-1:1992 & EN 61000-6-1:2001).
Tested & complies FCC Part 15 Subparts A & B. FDA CFR21 Part 11 Compliance
FDA CFR21 Part 11 Compliance
Designed for FDA CFR21 Part 11 Digital signatures
RoHS Compliance
RoHS Directive
The RoHS Directive stands for "the restriction of the use of certain hazardous substances in electrical and electronic equipment". This Directive bans the placing on the EU market of new electrical and electronic equipment containing more than agreed levels of lead, cadmium, mercury, hexavalent chromium, polybrominated biphenyl (PBB) and polybrominated diphenyl ether (PBDE) flame retardants.
QUESTIONS & ANSWERS
Ask a Question-
Is it calibrated?
It should work perfectly out of the box. It is not calibrated, but calibration can be purchased with the item, which can be found here on our website: https://techinstrument.com/products/utrix-16-logtag-multi-use-usb-pdf-temperature-recorder