TRIX-8 Multi use Temperature Data Recorder

$32.00 USD

TRIX-8

The LogTag® TRIX-8 data logger is a versatile, wide range, multi-trip Temperature Recorder, featuring high resolution temperature readings over a measurement range of -40°C to +85°C (-40°F to +185°F).

The LogTag® TRIX-8 calibrated temperature data logger is equipped with a unique external temperature sensor arrangement providing fast reaction time to temperature change. Protect the TRIX-8 from damage with a LogTag Protective Enclosure for your data logger.  Of course, we offer our temperature data logger calibration service for this and all of our data loggers.

  • PLEASE ALLOW UP TO 2 BUSINESS DAYS FOR CERTIFICATION COMPLETION, IF ORDERED.


IMPORTANT:  Each location (not each logger) requires a LogTag Cradle in order to program and retrieve data from the logger.  

Interface Cradle


Using the LogTag® Interface Cradle and the freely available companion software LogTag® Analyzer, the LogTag® TRIX-8 data logger is easily set-up for recording including delayed start, sampling interval, number of readings and configuration of conditions to activate the 'ALERT' indicator.

LogTag® Analyzer also provides facilities for charting, zooming, listing data statistics and allows exporting the data to other applications such as MS Excel once the data has been downloaded.

A real time clock provides date/time stamps for each temperature reading.

Push-to-start button with optional delay or a specific time & date. This portable temperature data logger comes with comprehensive customization options including alert settings, sample interval and trip duration. Complies with industry standards including EN12830. WHO PQS Prequalified under E006 devices : PQS Code E006/006

Highlights

  • Alert indicator – indicates if readings are outside of preset limits
  • OK indicator indicates if still recording and if readings within configured limits
  • Inspection mark in log by push button
  • Push Button Logging start
  • Rapid Download! Takes only seconds to download recordings
  • 'Pre-Start' logging - LogTag® can be configured to record even if it has not been started.
  • High performance at low cost
  • Credit card sized case – thin enough to be easily mailed “letter rate”.
  • Real time clock records time & temperature simultaneously
  • Easy to use LogTag®Analyzer software that runs on any PC configures LogTag® for recording then downloads resulting data for analysis. Data can also be exported to formats compatible with other applications such as Excel
  • Highly reliable and accurate
  • Re-calibration is possible to achieve higher accuracy possible

 

 

Specifications

Model Trix-8
Measurement Range -40°C ~ +85°C  (-40°F ~ +185°F)
Resolution < 0.1°C for -40°C ~ +40°C,
< 0.2°C for +40°C ~+80°C
Accuracy Better than ±0.5°C for -20°C~ +40°C.
Better than ±0.7°C for -20°C~ -30°C & +40°C~+60°C
Better than ±0.8°C for -30°C~ -40°C & +60°C~+80°C
Sensor Reaction Time Typically less than 5 minutes(T90) in moving air (1m/s).
Capacity 8032 readings (16K bytes memory)
Sampling Frequency Adjustable, 30 sec to several hours
Logging Start Options Push button start or specific date & time. Optional start delay of up to 18hours.
Recording Indication Flashing 'OK' indicator / flashing 'ALERT' indicator.
Download Time Typically with full memory (8000 readings) in less than 5 secs depending on computer or readout device used.
Environmental IP65 (roughly equivalent to NEMA 4)
Power Source 3V LiMg battery
Battery Life 2 ~ 3 years of normal use (based on 15 minute logging, download data monthly.
Size 86mm(H)x54.5mm(W)x8.6mm(T)
Weight 35g
Case Material Polycarbonate
Other Features • Logging start by push button or specific date/time start
• Optional clearing of alert indication by push button (places inspection make at same time).
• Download inspection mark recorded in LogTag memory every time LogTag is downloaded.
• 'Prestart' fail-safe logging (records temperature data even if not started)
• Low Battery indication in software.
• Comprehensive Alert indicator configuration.
• Calibration to achieve higher accuracy possible.

 

Calibration

In order to calibrate your LogTag Recorder, we use the LogTag Calibrate software. LogTag Recorders releases our LogTag Calibrate software only to selected users who have suitable equipment and credentials relating to standard practices of Temperature and/or Humidity calibration.

If you require calibration of your logger, please contact Tech Instrumentation at 303-841-7567 ext 2.


Calibrate after a battery change

Loggers that have had their case opened have potentially been exposed to electrostatic effects. In addition, units that had their battery replaced may not have been correctly hibernated or may have suffered from memory corruption if battery voltage was critically low during use.

They should therefore be validated for correct operation and performance after the battery change. The preferred way of doing this is by way of re-calibration, where ideally the product should be exposed to its full temperature measuring range.

 Additional Info

 Response Time for TRIX-8

The most common industry standard expression for temperature reading stabilization time is the ‘T90’. This is defined as the typical time taken of a given sensor in a given environment (e.g. moving air, still air, liquid etc.) to register 90% of an immediate step change in temperature.

The T90 of TRIX-8 is typically less than 5 minutes in moving air of 1m/s.

As the thermal step response is basically dependent on the latent heat of the two thermal masses involved (the Logger and the surrounding environment) and the rate of thermal transfer between them, the T90 is the same irrespective of the difference of temperatures though at extremes (water freezing and boiling points for example) the environment does not behave homogeneously thereby changing the rates of thermal conduction and therefore changing the resulting T90 value.

For example: in a situation of a 25°C step change say starting at 10°C and moving to 35°C in moving air of 1m/s, the TRIX-8 will typically register 90% of the step after less than 5 minutes – i.e. 0.9×25 = 22.5 -> 10+22.5 = registering 32.5°C within 5mins.

The T90 for a 15°C step in 1 m/s moving air (say 10°C to 25°C) will be the same. 0.9×15 = 13.5°C +10 = registering 23.5°C within 5mins.

T90 is different in slower moving air or still air, though air is never totally ‘still’ when a temperature difference exists as convection will take place.

 



Shipping & Delivery

We can ship to virtually any address in the world. Note that there are restrictions on some products, and some products cannot be shipped to international destinations.

Terms And Conditions

Making A Purchase

Making a purchase could not be easier. Just browse our store and add any items that you wish to buy into the shopping cart. After you have finished your selections, click on 'Checkout' and you will be asked for a few details that we need to be able to complete the order.


We accept credit cards and PayPal for payment. If you are shopping from outside of the USA, place your order and your credit card company will convert the transaction to your own currency. Please note that we can not ship to all international locations. If we are unable to ship to your location, we will advise you and your credit card will not be charged.

We accept Discover, Visa, Master Card and American Express. We do not charge for any item until it is ready to ship. Back-ordered items are not charged until they are shipped. You may send your credit card information via phone, fax, snail mail or securely over the Internet.

When confirmation of order is received, this is to indicate that we have received your order. It does not indicate that a contract exists between us. We will indicate acceptance of your order, and hence a contract between us, when we send you an invoice. We have included this term to protect us in the case that a mistake has been made in pricing, we have inadvertently underpriced goods, or we are no longer able to supply a particular product for some reason. In the case of a change of price, we will always contact you first to ensure that the price is acceptable.
 Credit Card Security
 When you place your order with us, your credit card number is encrypted for transmission to us using 128 bit encryption. Your credit card details are never by us since they are transmitted directly to the processing company.  As such, all refunds must be made to the same card used for the purchase.

Contacting Us
If you need to reach us, please email us using the link on the store page, alternatively, you can reach us at:

Tech Instrumentation Inc.
750 E Kiowa Ave.
PO Box 2029
Elizabeth, CO 80107-2029
USA

Phone: [800) 390-0004 Toll free
Phone: (303) 841-7567
Fax: (303) 840-8568

NOTE: We do not accept unsolicited advertisements at the facsimile number.


Email: info@techinstrument.com

Shipping And Handling
Shipping and handling charges are shown on a page just prior to your placing the order. We can ship to many parts of the world that are not allowed on the site. Please e-mail us for a shipping quote. Please call if you have any questions. 
PLEASE NOTE: If you return an item that you ordered with free shipping, we may, at our option deduct our shipping cost from your refund.

Delivery Schedule
We will normally ship your order to you within 1 to 3 business days. We will notify you via e-mail or telephone if we will be unable to ship within 3 business days. If your order is urgent, please let us know. We'll do our best to expedite urgent orders.
Back Orders
If your item is not in stock and we will not be able to ship within 10 business days, we will back order for you. You will always be emailed with the option to cancel your order if you would rather not wait.
Tax Charges
Orders made from within Colorado - sales tax will be applied. We do not currently charge taxes outside of Colorado, but some of our payment processors may charge sales tax and remit it directly to your state.
Guarantee
Most of our products come with a one year warranty. Most of the Cooper-Atkins thermometers which carry a five-year warranty. Most probes come with a one year warranty. Buyer assumes full responsibility in determining the suitability of these items for buyers intended use.
Privacy Policy
Tech Instrumentation will not disclose buyers' information to third parties other than when order details are processed as part of the order fulfillment.  Cookies are used on this shopping site to keep track of the contents of your shopping cart, to store delivery addresses if the address book is used and to store your details if you select the 'Remember Me' Option. They are also used after you have logged on as part of that process. You can turn off cookies within your browser by going to 'Tools | Internet Options | Privacy' and selecting to block cookies. If you turn off cookies, you will be unable to place orders or benefit from the other features that use cookies.

Returns Policy
Buyer assumes full responsibility in determining the suitability of these items for buyers intended use.  Most items come with a minimum one year warranty against defects in manufacturing.

CLAIMS FOR SHORTAGES OR ITEMS DAMAGED IN TRANSIT MUST BE SUBMITTED TO US WITHIN 3 DAYS OF RECEIPT OF THE PRODUCT. IF DAMAGE IS NOTED UPON DELIVERY, PLEASE NOTIFY CARRIER IMMEDIATELY AND BE SURE TO KEEP THE SHIPPING CONTAINER AND ALL PACKING MATERIAL. NO CLAIMS FOR SHORTAGES OR CARRIER DAMAGE WILL BE HONORED BEYOND 3 DAYS AFTER DELIVERY.

New and unopened items sold by Tech Instrumentation.com may be returned within 30 days of delivery. All items require prior authorization from Tech Instrumentation before they can be returned. We will also pay the return shipping costs if the return is a result of our error (e.g. you received an incorrect or defective item). New and unopened items being returned due to a customer error in ordering or handling will be subject to a restocking fee of 30% and must be returned freight prepaid.

All returns require a RGA number from Tech Instrumentation prior to being returned. Items must be returned postage paid to our facility in order to be considered for refund. Please call 303 841-7567 to obtain a RGA number if you need to return an item. Non defective returns must be in the original unopened packaging with all accessories to be eligible for a refund. Unopened items are subject to a 20% inspection and restocking charge. Non defective items that have been opened or are missing any parts including instructions are not eligible for return. Defective items will be repaired or replaced per mfg. policy. Shipping & Handling charges are not refundable. Service charges (repairs, certifications etc.) and special order items are never refundable. Non defective items must be returned within 30 days of the date we shipped to you in order to be eligible for refund. We pay shipping back you your location for all warranty returns. Items that have been damaged due to misuse or abuse, as judged by Tech Instrumentation are never eligible for a refund.

Refused/Undeliverable shipments: We will refund a minimum 75% of the cost of the product less shipping costs (both ways) for shipments that are returned to us as refused or undeliverable. Service charges (repairs, certifications etc.) are warranted but never refundable. Purchaser agrees to pay all charge-back and related costs incurred by us in the event a credit card charge-back is initiated by the purchaser for refused shipments. Please contact us directly and we will refund per our policies.  Shipping and handling charges are not refundable except in the event of an error on our part.

by LogTag
Type: Data Loggers

Customer Reviews

Based on 1 review
100%
(1)
0%
(0)
0%
(0)
0%
(0)
0%
(0)
A
Anna H./.R.G.(. (San Nicolás de los Garza, MX)
good seller

It is a Good product, I recommend it.

The LogTag® TRIX-8 data logger is a versatile, wide range, multi-trip Temperature Recorder, featuring high resolution temperature readings over a measurement range of -40°C to +85°C (-40°F to +185°F).

The LogTag® TRIX-8 calibrated temperature data logger is equipped with a unique external temperature sensor arrangement providing fast reaction time to temperature change. Protect the TRIX-8 from damage with a LogTag Protective Enclosure for your data logger.  Of course, we offer our temperature data logger calibration service for this and all of our data loggers.

  • PLEASE ALLOW UP TO 2 BUSINESS DAYS FOR CERTIFICATION COMPLETION, IF ORDERED.


IMPORTANT:  Each location (not each logger) requires a LogTag Cradle in order to program and retrieve data from the logger.  

Using the LogTag® Interface Cradle and the freely available companion software LogTag® Analyzer, the LogTag® TRIX-8 data logger is easily set-up for recording including delayed start, sampling interval, number of readings and configuration of conditions to activate the 'ALERT' indicator.

LogTag® Analyzer also provides facilities for charting, zooming, listing data statistics and allows exporting the data to other applications such as MS Excel once the data has been downloaded.

A real time clock provides date/time stamps for each temperature reading.

Push-to-start button with optional delay or a specific time & date. This portable temperature data logger comes with comprehensive customization options including alert settings, sample interval and trip duration. Complies with industry standards including EN12830. WHO PQS Prequalified under E006 devices : PQS Code E006/006

  • Alert indicator – indicates if readings are outside of preset limits
  • OK indicator indicates if still recording and if readings within configured limits
  • Inspection mark in log by push button
  • Push Button Logging start
  • Rapid Download! Takes only seconds to download recordings
  • 'Pre-Start' logging - LogTag® can be configured to record even if it has not been started.
  • High performance at low cost
  • Credit card sized case – thin enough to be easily mailed “letter rate”.
  • Real time clock records time & temperature simultaneously
  • Easy to use LogTag®Analyzer software that runs on any PC configures LogTag® for recording then downloads resulting data for analysis. Data can also be exported to formats compatible with other applications such as Excel
  • Highly reliable and accurate
  • Re-calibration is possible to achieve higher accuracy possible

 

 

Model Trix-8
Measurement Range -40°C ~ +85°C  (-40°F ~ +185°F)
Resolution < 0.1°C for -40°C ~ +40°C,
< 0.2°C for +40°C ~+80°C
Accuracy Better than ±0.5°C for -20°C~ +40°C.
Better than ±0.7°C for -20°C~ -30°C & +40°C~+60°C
Better than ±0.8°C for -30°C~ -40°C & +60°C~+80°C
Sensor Reaction Time Typically less than 5 minutes(T90) in moving air (1m/s).
Capacity 8032 readings (16K bytes memory)
Sampling Frequency Adjustable, 30 sec to several hours
Logging Start Options Push button start or specific date & time. Optional start delay of up to 18hours.
Recording Indication Flashing 'OK' indicator / flashing 'ALERT' indicator.
Download Time Typically with full memory (8000 readings) in less than 5 secs depending on computer or readout device used.
Environmental IP65 (roughly equivalent to NEMA 4)
Power Source 3V LiMg battery
Battery Life 2 ~ 3 years of normal use (based on 15 minute logging, download data monthly.
Size 86mm(H)x54.5mm(W)x8.6mm(T)
Weight 35g
Case Material Polycarbonate
Other Features • Logging start by push button or specific date/time start
• Optional clearing of alert indication by push button (places inspection make at same time).
• Download inspection mark recorded in LogTag memory every time LogTag is downloaded.
• 'Prestart' fail-safe logging (records temperature data even if not started)
• Low Battery indication in software.
• Comprehensive Alert indicator configuration.
• Calibration to achieve higher accuracy possible.

 

In order to calibrate your LogTag Recorder, we use the LogTag Calibrate software. LogTag Recorders releases our LogTag Calibrate software only to selected users who have suitable equipment and credentials relating to standard practices of Temperature and/or Humidity calibration.

If you require calibration of your logger, please contact Tech Instrumentation at 303-841-7567 ext 2.


Calibrate after a battery change

Loggers that have had their case opened have potentially been exposed to electrostatic effects. In addition, units that had their battery replaced may not have been correctly hibernated or may have suffered from memory corruption if battery voltage was critically low during use.

They should therefore be validated for correct operation and performance after the battery change. The preferred way of doing this is by way of re-calibration, where ideally the product should be exposed to its full temperature measuring range.

 Response Time for TRIX-8

The most common industry standard expression for temperature reading stabilization time is the ‘T90’. This is defined as the typical time taken of a given sensor in a given environment (e.g. moving air, still air, liquid etc.) to register 90% of an immediate step change in temperature.

The T90 of TRIX-8 is typically less than 5 minutes in moving air of 1m/s.

As the thermal step response is basically dependent on the latent heat of the two thermal masses involved (the Logger and the surrounding environment) and the rate of thermal transfer between them, the T90 is the same irrespective of the difference of temperatures though at extremes (water freezing and boiling points for example) the environment does not behave homogeneously thereby changing the rates of thermal conduction and therefore changing the resulting T90 value.

For example: in a situation of a 25°C step change say starting at 10°C and moving to 35°C in moving air of 1m/s, the TRIX-8 will typically register 90% of the step after less than 5 minutes – i.e. 0.9×25 = 22.5 -> 10+22.5 = registering 32.5°C within 5mins.

The T90 for a 15°C step in 1 m/s moving air (say 10°C to 25°C) will be the same. 0.9×15 = 13.5°C +10 = registering 23.5°C within 5mins.

T90 is different in slower moving air or still air, though air is never totally ‘still’ when a temperature difference exists as convection will take place.